Table 2

Procedural characteristics

CategorymITT analysis set n=100Safety analysis set n=102
Type of sedation used, n/N (%)
General anesthesia59/100 (59.0%)59/102 (57.8%)
Local anesthesia7/100 (7.0%)7/102 (6.9%)
Conscious sedation34/100 (34.0%)36/102 (35.3%)
Femoral arterial puncture, n/N (%)100/100 (100.0%)102/102 (100.0%)
Time to arterial puncture since stroke onset, n9496
Median (IQR) (minutes)268.5 (183.0–389.0)268.5 (183.5–388.0)
Mean±SD (minutes)377.1±357.7374.0±354.6
Time to first aspiration attempt since stroke onset, n9494
Median (IQR) (minutes)281.0 (199.0–400.0)281.0 (199.0–400.0)
Mean±SD (minutes)396.1±357.0396.1±357.0
Time to first aspiration attempt since stroke onset by subgroup, n/N (%)
≤8 hours74/94 (78.7%)74/94 (78.7%)
>8 hours20/94 (21.3%)20/94 (21.3%)
Type of post-thrombectomy intervention if used, n/N (%)6/100 (6.0%)6/102 (5.9%)
Proximal lesion stenting1/6 (16.7%)1/6 (16.7%)
Target lesion stenting5/6 (83.3%)5/6 (83.3%)
Target lesion angioplasty5/6 (83.3%)5/6 (83.3%)
Use of stent retrievers, n/N (%)
At least once35/100 (35.0%)35/102 (34.3%)
First use of stent retriever within first 3 procedural passes for vessels >2.0 mm6/100 (6.0%)6/102 (5.9%)
First use of stent retriever within first 3 procedural passes for vessels <2.0 mm14/100 (14.0%)14/102 (13.7%)
First use of stent retriever after 3 passes15/100 (15.0%)15/102 (14.7%)
None65/100 (65.0%)67/102 (65.7%)
Use of BGC, n/N (%)5/100 (5.0%)5/102 (4.9%)
Use of microcatheter, n/N (%)84/100 (84.0%)85/102 (83.3%)
Total number of EMBOVAC passes, n/N (%)
Zero passes0/100 (0.0%)2/102 (2.0%)
One pass59/100 (59.0%)59/102 (57.8%)
Two passes15/100 (15.0%)15/102 (14.7%)
Three passes11/100 (11.0%)11/102 (10.8%)
Four passes4/100 (4.0%)4/102 (3.9%)
Five passes9/100 (9.0%)9/102 (8.8%)
Six passes1/100 (1.0%)1/102 (1.0%)
Eleven passes1/100 (1.0%)1/102 (1.0%)
Total number of EMBOVAC passes, mean±SD2.0±1.62.0±1.6
Total number of non-EMBOVAC passes, mean±SD0.4±1.10.5±1.2
Total number of passes (all devices), mean±SD2.4±2.22.4±2.2
Rescue therapy during the first three passes of EMBOVAC10/100 (10.0%)11/102 (10.8%)
Any change in frontline device therapy to remove the target occlusion in a vessel of at least 2.0 mm in size7/10 (70.0%)7/11 (63.6%)
Pass 11/7 (14.3%)1/7 (14.3%)
Pass 23/7 (42.9%)3/7 (42.9%)
Pass 34/7 (57.1%)4/7 (57.1%)
Use of intracranial lesion stenting during procedure5/10 (50.0%)5/11 (45.5%)
Use of intra-arterial thrombolytic agent during the procedure0/10 (0.0%)0/11 (0.0%)
  • BGC, balloon guide catheter; mITT, modified intent-to-treat.